The R&S®NRP-Z51 thermal power sensor with its outstanding accuracy is suitable for very demanding reference applications independent of the signal modulation type.
Key Facts• Measurement range: -35 dBm to + 20 dBm
• Frequency range: DC to 18 GHz
• N connector
• Operation with R&S®NRP2 base unit or via USB on a Laptop/PC
• Outstanding linearity of 0.007 dB (0.16%)
• Industry-proven DC-coupled thermoelectric test cell
• Highly accurate continuous average power measurements
• Sophisticated connector design
• Excellent impedance matching
Supported DevicesThe R&S®NRP-Z51 sensor can be connected to latest-generation of signal generators, spectrum analyzers and network analyzers. This allows you to directly display the average power measured with the sensor without the need for an R&S®NRP2 base unit or PC. This also provides a simple solution for automatic level correction in a signal generator. The spectrum analyzers use the measured power as an absolute reference for increasing the measurement accuracy.
Especificações
Thermal Power Sensor
Frequency range
| Power measurement range; maximum input power
| Impedance matching (SWR)
|
| Rise time, video BW
| Uncertainty for power measurements at +20 °C to +25 °C (1) absolute (in dB) (2) relative (in dB) |
DC to 18 GHz
| -35 dBm to +20 dBm (300 nW to 100 mW); +25 dBm (AVG), +40 dBm (PK, 1 µs)
| DC to 100 MHz >100 MHz to 2.4 GHz >2.4 GHz to 12.4 GHz >12.4 GHz to 18.0 GHz
| < 1.03 < 1.06 < 1.13 < 1.16
| -
| (1) 0.040 to 0.082 (2) 0.010 |