HP - Agilent |
Key specifications
Agilent 4284A precision LCR meter
Test frequency 20 Hz to 1 MHz over 8610 selectable frequencies
Measurement range1
|Z|, R, X :0.01 m?to 99.9999 M?
|Y|, G,B :0.01 nS to 99.9999 S
C :0.01 fF to 9.99999 F
L :0.01 nH to 99.9999 kH
D :0.000001 to 9.99999
Q :0.01 to 99999.9
Basic accuracy |Z| , C and L:0.05% D:0.0005
Test signal level range
Voltage 5 mVrms to 2 Vrms
Current 50 µArms to 20 mArms
Constant test signal level range
Voltage 10 mVrms to 1 Vrms
Current 100 µArms to 10 mArms
Measurement time2
39 ms/190 ms/830 ms at 1 kHz
Acessorios:
Agilent 16334A
Agilent 16047A
Agilent 16048C
Agilent 16034G
Agilent 16065A external voltage bias fixture